PMDO-700 Series Online PPB Low-Level DO Dissolved Oxygen Analyzer Continuously Monitor Test Pure Water Trace
Product Description
PMDO-700 Series On-line Trace Dissolved Oxygen Analyzer
1.1 Measurement Principle
The trace dissolved oxygen analyzer is a highly sensitive on-line analytical instrument. Composed of a main unit, sensor, flow cell, and cable, it features high sensitivity, fast response, and user-friendly operation, specifically designed for measuring trace-level (ppb-level) oxygen concentration in gas-water/gas media.
Oxygen measurement adopts the electrochemical polarography method. During measurement, the oxygen sensor undergoes a redox reaction to generate a diffusion current, whose magnitude is proportional to the oxygen concentration in the sample.
1.2 Functional Features
· High-resolution color touch screen: 800×480 resolution, Chinese menu, easy operation
· Real-time curve display: Simultaneous digital and curve presentation for intuitive results
⭐ Multi-channel setting: Supports 4 oxygen sensors for cost efficiency
· Convenient installation & maintenance: Direct fixed mounting (no base needed); rear wiring for reliability
⭐ Plug-in sensor design: Facilitates easy maintenance and replacement
· High precision & sensitivity: Detection limit down to 0.1μg/L or 0.01μg/L; resolution 0.1μg/L or 0.01μg/L
⭐ Fast response: Typically 3 minutes from air to μg/L-level concentration
· Long maintenance cycle & low consumable cost: Electrode maintenance every 6–8 months under normal use
⭐ Large data storage: Holds 1.5 million records with historical curve query function
· 4–20mA current output and RS485 communication
1.3 Typical Application Scenarios
· Trace dissolved oxygen analysis for deoxygenated water in power plants and waste heat boilers
· Trace oxygen monitoring in ultra-pure water processes for semiconductor chip packaging and high-end PCB manufacturing
1.4 Technical Parameters
Parameter | Specification |
Display | 800×480 color touch screen, Chinese menu |
Channels | Up to 4 dissolved oxygen channels |
Measurement Range | 0.01μg/L–20.00mg/L |
Resolution | 0.1μg/L or 0.01μg/L (optional) |
Measurement Error | ±1.5%FS or 1.0μg/L (whichever is larger) |
Response Time (T90) | <30s |
Medium Temperature | 0–50℃ |
Temperature Resolution | 0.1℃ |
Sample Flow Rate | 100–400mL/min |
Working Environment | Temperature: 0–50℃; RH: ≤90% |
Output Current | 0/4–20mA |
Current Error | ±0.01mA |
Max. Load Resistance | 800Ω (at 20mA) |
Alarm Outputs | 2-way alarm relays |
Power Supply | AC 220V / DC 24V |
Data Storage | 1.5 million records |
Communication | RS485 |
Weight | 800g |
Installation | Cutout/wall-mounted/pipeline |
Cutout Size | 138mm×138mm |