A reliable ISO direct manufacturer.                                                                                                         Please contact  info@probesti.com

PRODUCTS DETAIL

You are here: Home » Products » Sensor and Controller » China Wholesale ISO Manufacturer PEA Method Low-range Turbidity Analyzer Manufacturer Digital Turbidity Meter Drinking Water Test

SITE SEARCH

PRODUCT CATEGORY

GET IN TOUCH

 Email: 
info@probesti.com
zhongweihua@probest.cn 
  Tel: +8613950412146
  Skype: zhongweihua 

loading

Share to:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
whatsapp sharing button
kakao sharing button
snapchat sharing button
sharethis sharing button

China Wholesale ISO Manufacturer PEA Method Low-range Turbidity Analyzer Manufacturer Digital Turbidity Meter Drinking Water Test

 Measurement Principle
Incident light of PLTU‐700 Low‐Range Sensor is produced by the tungsten light whose color temperature’s absolute temperature is between 2200 degrees to 3000 degrees.Guiding the parallel light assembled by the tungsten light down into the water sample of the sensor,the light is scattered by suspended particles in water samples,and the scattered light that is with a ninety degree angle to the angle of incidence is received by the silicon photocell receiver in the water sample,measuring the turbidity by calculating the relationship with 90‐degree scattered light and the incident beam.
Availability:
Quantity:
  • PLTU-700

  • PROBEST

  • 90268090

PLTU-700 China Wholesale ISO Manufacturer PEA Method Low-range Turbidity Analyzer Manufacturer Digital Turbidity Meter Drinking Water Test

n  Measurement Principle

Incident light of PLTU700 LowRange Sensor is produced by the tungsten light whose color temperature’s absolute temperature is between 2200 degrees to 3000 degrees.Guiding the parallel light assembled by the tungsten light down into the water sample of the sensor,the light is scattered by suspended particles in water samples,and the scattered light that is with a ninety degree angle to the angle of incidence is received by the silicon photocell receiver in the water sample,measuring the turbidity by calculating the relationship with 90degree scattered light and the incident beam.

n  Features

u  A scattering turbidity meter with continuous readings, designed for low‐range turbidity monitoring.

u  Stable data and good repeatability.

u  Simple clean and maintenance.

u  Plug-and-play sensors,quick and easy installation.

n  Applications

u  Turbidity online monitoring of filtration of water plant, product Water and direct drinking water system.

u  Turbidity online monitoring of recirculating cooling water and filter water in all kinds of industrial production and water reuse system.

 

n  Technical Specifications

 

Measurement range

0.001-100 NTU

Measurement   Accuracy

±2% of the reading or   ±0.015NTU when the range is

0.001-40NTU,whichever   is greater;±5% of the reading

when the range is   40-100NTU

Water flow

300ml/min≤X≤700ml/min

Calibration

Standard solution   calibration,water ample calibration,zero

calibration

Pipe Fitting

Inlet1/4NPTOutlet1/2NPT

Main   material

BodyABS   + SUS316L Seal: Butyl Cyanide RubberCablePUR

Power supply

AC85-500VAC (50/60HZ)  DC9~36VDC

Output

3-way 4-20mA

Relay

Three-way relays, programmed response parameter

and response value.

Communication protocol

MODBUS  RS485

Storage temperature

-15-60

Measuring   temperature

0-45Non-freezing

Dimension

Sensor310 *210*410mm(L*W*H)

Transmitter145*125*162mm(L*W*H)

Weight

Sensor2.1KGTransmitter1.35KG

Protective rate

SensorIP65(Indoor)TransmitterIP65/   NEMA4X

Cable   length

Standard: 3m,not recommend to   extend.