  • Q Q7: What practical benefits does PROBEST’s dual-standard calibration design offer users?

    A Q7: What practical benefits does PROBEST’s dual-standard calibration design offer users?

    Answer:
    High versatility: Compatible with two mainstream commercial pH buffers on the market, more flexible for buffer purchasing and cost saving;
    Higher testing precision: Match calibration buffer with sample pH property to minimize measuring error;
    Multi-industry availability: One PROBEST pH meter meets testing needs of wastewater, food, chemical and pure water industries.
  • Q Q5: How should the PROBEST electrode be prepared before each calibration with a new standard buffer solution?

    A Q5: How should the PROBEST electrode be prepared before each calibration with a new standard buffer solution?

    Answer: Rinse electrode fully with clean water and blot residual water with lint-free tissue before next buffer calibration to avoid cross-contamination and inaccurate calibration.
  • Q Q4: When calibrating the 6.86, 4.0, and 9.18 standard solutions, what mV reading is considered within the normal range of PROBEST pH sensor?

    A Q4: When calibrating the 6.86, 4.0, and 9.18 standard solutions, what mV reading is considered within the normal range of PROBEST pH sensor?

    Answer:
    • pH6.86 buffer: mV reading within ±15 mV is normal;
    • pH4.0 buffer: Normal reading around 170 mV±20mV;
    • pH9.18 buffer: Normal reading around 120 mV±20mV.
    Abnormal mV values mean expired buffer, dirty or damaged pH electrode.