PTU-800
PROBEST
90268090
n Features
u The sensor adopts dual-beam infrared scattered light
photometer detection technology and has good repeatability and stability.
u Use ISO7027 Standard Method (Infrared light scattering technique) to eliminate effects of sample color.
u Built-in self-diagnosis function to ensure the accuracy of data.
u Optional cleaning brush of automatic cleaning function to reduce the maintenance of sensor.
u Digital sensor,high anti-jamming capacity and far transmission distance.
u Standard digital signal output,can achieve integration and networking with other equipment without controller.
u Plug-and-play sensors,quick and easy installation.
n Applications
u Turbidity online monitoring of water inlet, settling basin of water plant.
u Turbidity online monitoring of sewage plant, industrial process water and waste water treatment.
n Technical Specifications
Measurement range | Turbidity:0.01-100 NTU,0.01-4000 NTU |
Measurement accuracy | Less than±2% of the measured reading or ±0.1NTU,whichever is greater |
Pressure range | ≤0.4Mpa |
Flow rate | ≤2.5m/s、8.2ft/s |
Calibration | Sample calibration,Slope calibration |
Main material | Body:SUS316L(fresh water),Titanium alloy (marine); Cover:POM;Cable:PUR |
Power supply | AC:85-500VAC (50/60HZ) DC:9~36VDC |
Output | 3-way 4-20mA |
Relay | Three-way relays, programmed response parameter and response value. |
Communication protocol | MODBUS RS485 |
Storage temperature | -15-50℃ |
Measuring temperature | 0-45℃(Non-freezing) |
Dimension | Sensor:Dia 60mm*L 256mm Transmitter:145*125*162mm(L*W*H) |
Weight | Sensor:1.65KG;Transmitter:1.35KG |
Protective rate | Sensor:IP68/NEMA6P;Transmitter:IP65/ NEMA4X |
Cable length | Standard:10m,the maximum may be extended to 100m |
n Features
u The sensor adopts dual-beam infrared scattered light
photometer detection technology and has good repeatability and stability.
u Use ISO7027 Standard Method (Infrared light scattering technique) to eliminate effects of sample color.
u Built-in self-diagnosis function to ensure the accuracy of data.
u Optional cleaning brush of automatic cleaning function to reduce the maintenance of sensor.
u Digital sensor,high anti-jamming capacity and far transmission distance.
u Standard digital signal output,can achieve integration and networking with other equipment without controller.
u Plug-and-play sensors,quick and easy installation.
n Applications
u Turbidity online monitoring of water inlet, settling basin of water plant.
u Turbidity online monitoring of sewage plant, industrial process water and waste water treatment.
n Technical Specifications
Measurement range | Turbidity:0.01-100 NTU,0.01-4000 NTU |
Measurement accuracy | Less than±2% of the measured reading or ±0.1NTU,whichever is greater |
Pressure range | ≤0.4Mpa |
Flow rate | ≤2.5m/s、8.2ft/s |
Calibration | Sample calibration,Slope calibration |
Main material | Body:SUS316L(fresh water),Titanium alloy (marine); Cover:POM;Cable:PUR |
Power supply | AC:85-500VAC (50/60HZ) DC:9~36VDC |
Output | 3-way 4-20mA |
Relay | Three-way relays, programmed response parameter and response value. |
Communication protocol | MODBUS RS485 |
Storage temperature | -15-50℃ |
Measuring temperature | 0-45℃(Non-freezing) |
Dimension | Sensor:Dia 60mm*L 256mm Transmitter:145*125*162mm(L*W*H) |
Weight | Sensor:1.65KG;Transmitter:1.35KG |
Protective rate | Sensor:IP68/NEMA6P;Transmitter:IP65/ NEMA4X |
Cable length | Standard:10m,the maximum may be extended to 100m |